Recalled ICD leads fail in women, youths most often

Heart Disease newsJan 31, 2011

The recalled Sprint Fidelis implantable cardioverter-defibrillator (ICD) leads (Medtronic) failed more often in younger patients, women, and individuals with hereditary heart disease, according to a multicenter study published online Jan. 17 in Circulation. The researchers found that lead failure was not associated with death or serious injuries.

However, about half of the patients whose leads fractured experienced painful inappropriate shocks, according to lead author Robert G. Hauser, MD, of the Minneapolis Heart Institute® at Abbott Northwestern Hospital. In a previous study, published in HeartRhythm in 2010, Hauser and colleagues found these inappropriate shocks could be mitigated by a software program available in Medtronic ICDs.

The manufacturer voluntarily removed Sprint Fidelis leads from the market in 2007 after 268,000 were implanted globally. Currently, more than 100,000 patients have Fidelis leads. 

In this observational study, 1,023 Fidelis and 1,668 Quattro leads were implanted and followed at three centers (Minneapolis Heart Institute® in Minneapolis, Mayo Clinic in Rochester, Minn., and Beth Israel Deaconess Medical Center in Boston) from November 2001 to January 2009

“We were particularly interested in whether Fidelis leads were continuing to fail, in whom they were failing and what the implications might be for those patients,” Hauser says. “We found that patients who were young and active with relatively normal pump function were at highest risk, as well as women compared with men.”

The researchers found the failure rate for Fidelis leads was 2.81 percent per year, compared with 0.43 percent per year for Quattro Secure leads, also manufactured by Medtronic.. The authors described the Quattro failure rate as being “well below the suggested benchmark of 0.6 percent per year. The survival of Fidelis leads at four years was 87 percent, compared with 98.7 percent for Quattro leads

These findings could have significant implications for the management of patients who have Fidelis leads. While Fidelis continue to fail, Hauser and his colleagues conclude that there is currently no indication for arbitrarily replacing leads in patients who were indicated for ICDs, particularly due to the potential risk with the replacement procedures.

“Only in the hands of experienced operators should Fidelis lead replacement even be considered in these younger patients or women, who are expected to live with an ICD for a number of years,” Hauser stresses.

###

This study was funded by research grants from the Minneapolis Heart Institute Foundation and the Abbott Northwestern Hospital Foundation.

Minneapolis Heart Institute®

The Minneapolis Heart Institute® is recognized internationally as one of the world’s leading providers of heart and vascular care. This state-of-the-art facility combines the finest in personalized patient care with sophisticated technology in a unique, family-oriented environment. The Institute’s programs, a number of which are conducted in conjunction with Abbott Northwestern Hospital, address the full range of heart and vascular health needs: prevention, diagnosis, treatment and rehabilitation.

Minneapolis Heart Institute Foundation

The Minneapolis Heart Institute Foundation is dedicated to creating a world without heart disease through groundbreaking clinical research and innovative education programs. MHIF’s mission is to promote and improve cardiovascular health, quality of life and longevity for all.

###

Contact: Meghan Bethke

612-863-5410
Minneapolis Heart Institute Foundation

Provided by ArmMed Media

Email this to a friend Bookmark this! Printable Version

RELATED STORIES:


 Comments [ + Post Your Own

Now you're in the public comment zone. What follows is not Armenian Medical Network's stuff; it comes from other people and we don't vouch for it. A reminder: By using this Web site you agree to accept our Terms of Service. Click here to read the Rules of Engagement.

There are no comments for this entry yet. [ + Comment here + ]




We are pleased to let readers post comments about an article. Please increase the credibility of your post by including your full name and email.

All comments are reviewed by our editors before they are posted on the site. Just keep it clean, kids.

Name:

Email:

Location:

URL:

Remember my personal information

Notify me of follow-up comments?

Please enter the word you see in the image below:


   [advanced search]   
What is Levonorgestrel Emergency contraceptive Kit. Levonorgestrel can prevent ovulation and pregnancy after unprotected sex


Health Centers

  Heart Attack

  Overview

  Causes

  Risk Factors

  Signs & Symptoms

  Diagnosis and Tests

  Treatment

  Prevention

  Follow-up

  Summary

  FAQ

  Conditions

  Angina

  Mitral stenosis

  Atrial Fibrillation

  Chest Pain

  Heart Failure

  Endocarditis

  Arrhythmias

  Atherosclerosis

  Heart disease Risk Factors

  Heart attack

  Coronary artery disease

  Coronary heart disease

  Congenital heart disease

  First aid - cardiac arrest

  Heart Surgery

  Myocardial Infarction

  Brady-tachycardia syndrome

  Anatomy of the Heart

» » »

Health Centers





Diabetes









Health news
  


Health Encyclopedia

Diseases & Conditions

Drugs & Medications

Health Tools

Health Tools



   Health newsletter

  





   Medical Links



   RSS/XML News Feed



   Feedback






Add to Google Reader or Homepage
Heart Diseases News, Headlines and Latest Stories on Health.am
Add to My AOL





Human Rights in Patient Care - Practitioner Guide